Part Number Hot Search : 
CAT65067 AX797HE 4VHC1G GH60N60 2SC4404 1602G 74LVC IG0512SA
Product Description
Full Text Search
 

To Download E6833A Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 Agilent E6833A cdma2000/1xEV-DO Calibration Application
For the E6601A Wireless Communications Test Set
Data Sheet
The next generation of mobile phone manufacturing test.
The Agilent E6601A is the new, one-box test set that expedites mobile phone calibration. Combining industry-leading measurement speed, selectable formats, flexible licensing, and an integrated open Windows(R) XP PC, the E6601A helps you achieve the lowest cost of test in mobile phone manufacturing. The E6601A and its available technology-specific software applications deliver industry leading measurement speed and accuracy for your mobile phone test needs. The Agilent E6833A cdma2000/1xEV-DO Calibration Application is a non-signaling application optimized for cdma2000(R)/1xEV-DO mobile phone calibration. The E6833A also provides a choice of perpetual, transportable, or term licenses for maximizing asset utilization and your cost per test. The test set and its applications use emerging advancements such as fast device tune features to deliver fast, high-quality calibration. Because it's Agilent, you can be confident--it operates using trusted, proven measurement methodology that ensures measurement integrity that is never compromised.
E6601A/E6890A Features and General Specifications
* * * * * * * * * * * CW, AM, FM, DSB-SC-SC source modulation RF analyzer Spectrum monitor Transmitter power measurement Power versus time measurement Frequency error measurement Optional IQ capture waveform sampling Internal OCXO timebase Built-in open Windows XP PC Run test programs with internal or external PC GPIB, USB, and LAN connectivity and control
E6833A Features
Source modulation * CW, amplitude, frequency, IS-95, cdma2000, and 1xEV-DO Power measurements * Mean power * Root-raised cosine power Power measurements * Channel power * Dynamic power * Fast device tune Spectral measurements * Tx spurious emissions * Occupied bandwidth * Spectrum monitor Modulation quality measurements * IS-95, cdma2000 RC1 rho
www..com
Technical Specifications
These specifications apply to an E6601A mainframe and the E6833A cdma2000/1xEVDO Calibration Application firmware revision A.03 or higher. Only feature additions beyond the E6890A General Purpose Application are included in this document. Specifications describe the test set's warranted performance and are valid for the unit's operation within 10 C of the last self alignment. All specifications are valid after a 30-minute warm-up period of continuous operation with valid self alignment. If the instrument has been off for longer than 48 hours, a 48-hour warm-up period followed by a self alignment is required. Supplemental characteristics are intended to provide typical, but non-warranted, performance parameters that may be useful in applying the instrument. These characteristics are shown in italics and labeled as "typical." All units shipped from the factory meet these typical numbers at +25 C ambient temperature without including measurement uncertainty.
RF Generator
Frequency
W-CDMA cellular bands US cellular band US PCS band Korean PCS band Japan CDMA band 1-799, 991-1023, 1024-1323 (860.04-893.97 MHz) 0-1199 (1930-1990 MHz) 0-599 (1840-1870 MHz) 1-799, 801-1039, 1041-1199, 1201-1600 (832.0125-869.9875 MHz) 0-1199 (2110-2169.950 MHz) 1-400, 472-871, 1039-1473, 1792-2016 (approx 420-494 MHz) 0-719, 720-919 (851-869, 935-940 MHz) 0-1299 (1930-1995 MHz) 0-899 (2110-2155 MHz)
IMT-2000 band NMT-450 band Secondary 800 band US PCS 1900 band AWS band
Amplitude
Output port control User control of RF source routing to either the RF in/out port or the RF out only port -115 dBm/1.23 MHz to -15 dBm/1.23 MHz -115 dBm/1.23 MHz to -5 dBm/1.23 MHz
Output level range RF in/out port RF out only port
Composite absolute output level accuracy (< 10 C and < 24 hours from the last alignment, -108 to -15 dBm/1.23 MHz) RF in/out port RF out only port < 1.0 dB, typically 0.65 dB < 1.0 dB, typically 0.65 dB
Modulation
Waveforms Spectrally correct IS-95, cdma2000 RC3, 1xEV-DO waveforms Typically > 0.99 Typically < 3%
Modulation quality Residual rho Residual EVM
www..com
2
RF Analyzer
Frequency
Frequency range in ch annel numbers US cellular band 1-799, 991-1023, 1024-1323 (815.04-848.97 MHz) US PCS band Korean PCS band Japan CDMA band 0-1199 (1850-1910 MHz) 0-599 (1750-1780 MHz) 1-799, 801-1039, 1041-1199, 1201-1600 (887.0125-924.9875 MHz) 0-1199 (1920-1979.950 MHz) 1-400, 472-871, 1039-1473, 1792-2016 (approx 410-484 MHz) 0-719, 720-919 (806-824, 896-901 MHz) 0-1299 (1850-1915 MHz) 0-899 (1710-1755 MHz)
Reverse Traffic Channel Suite
Triggering this measurement suite results in the capture of one block of samples from which is derived the channel power, occupied bandwidth, and Tx spurious emissions parameters. Trigger setup Sources External, immediate, RF generator ARB
Channel power measurement
Measurement method Measures the total power in a 1.23 MHz bandwidth centered on the reverse channel center frequency 1.25 ms for cdma2000 or 1.67 ms for 1xEV-DO -61 dBm/1.23 MHz to +28 dBm/1.23 MHz < 0.7 dB, typically < 0.2 dB Typically < 0.05 dB
IMT-2000 band NMT-450 band Secondary 800 band US PCS 1900 band AWS band
Measurement interval Measurement range Measurement accuracy -61 to +28 dBm/1.23 MHz Measurement repeatability
Amplitude
Input level range Channel power Peak input power Self alignment validity -61 dBm/1.23 MHz to +28 dBm/1.23 MHz -61 dBm/1.23 MHz to +37 dBm/1.23 MHz < 10 C change and < 30 days from last self alignment
Modulation quality measurement
Input frequency ranges Measurement chip rate Measurement method Input level range Same as RF analyzer 1.2288 Mcps IS-95 or cdma2000 RC1: single code rho Average power -25 to +37 dBm/ 1.23 MHz
Modulation quality measurement range (for signals with < 1 kHz frequency error) IS-95, RC1 Measurement interval 0.40 to 1.00 rho 1.042 ms (5 Walsh symbols)
Modulation quality measurement accuracy IS-95, RC1 rho < 0.003 + residual error for 0.8 < rho < 1.0 Modulation quality measurement residuals rho > 0.999 Frequency error Measurement results 2 Hz plus timebase error rho, frequency error, phase error, magnitude error, carrier feedthrough
www..com
3
Occupied bandwidth
Measurement method Measures the bandwidth that contains 99.0% of the total integrated power of the transmitted signal centered on the selected reverse channel +5 to +28 dBm/1.23 MHz 60 kHz for 99.0% total integrated power
Dynamic Power Measurement
Measurement method Captures a user-defined trace consisting of 5, 10, or 20 ms duration power steps with user-defined step size produced by a test mode in the device under test. Measures the total power in each step period in a 1.23 MHz bandwidth centered on the reverse channel center frequency -61 to +28 dBm/1.23 MHz With 4 dB crest < 2 dB of expected power
Input level range Measurement accuracy
Tx spurious emissions
Measurement method Measures the active carrier power in a 1.23 MHz bandwidth, then measures the power in a 30 kHz bandwidth at two offsets above and below the active carrier and displays the ratio of the offset powers to the active carrier power in dBc +0 to +30 dBm/1.23 MHz 885 kHz, 1.98 MHz 1.25 MHz, 1.98 MHz 1.23 MHz flat channel filter 30 kHz synchronously tuned, five-pole filter with approximately Gaussian shape < 0.35 dB, typically < 0.15 dB < 0.6 dB, typically < 0.25 dB < -73 dBc/30 kHz BW < -75 dBc/30 kHz BW
Input level range Amplitude capture range
Measurement accuracy1 (within +2 to -45 dB of expected power) -61 to +28 dBm/1.23 MHz < 0.7 dB, typically < 0.2 dB Measurement accuracy1 (within 45 dB of expected power with < 48 hours warm-up before self alignment initiation) -59 to +28 dBm/1.23 MHz < 0.7 dB, typically < 0.3 dB -65 to < -59 dBm/1.23 MHz < 0.8 dB, typically < 05 dB Measurement repeatability Typically < 0.05 dB for returning to the same level and frequency with no temperature change and insignificant time change Sources external, fall, rise
User specified maximum power Trigger pulse Step size
Input level range Measurement offsets Frequencies < 1000 MHz Frequencies > 1000 MHz Measurement bandwidth Carrier Offsets
Trigger setup
Measurement accuracy 885 kHz, 1.25 MHz 1.98 MHz Residual noise floor 885 kHz, 1.25 MHz 1.98 MHz offsets Mobile pass/fail limits Auto mode Frequencies < 1000 MHz
Step duration
-42 dBc/30 kHz for 885 kHz offsets -54 dBc/30 kHz for 1.98 MHz offsets
Frequencies > 1000 MHz
-42 dBc/30 kHz for 1.25 MHz offsets -50 dBc/30 kHz for 1.98 MHz offsets
Manual mode
Settable from -10 to -65 dBc with 0.01 dB resolution
www..com
1. Additional accuracy error when using RF OUT ONLY port is < 0.1 dB.
4
Fast Device Tune Measurement
Measurement method Allows simultaneous calibration of a device's transmitter output power and receiver input level across level and frequency in a single sweep (per frequency band). The device must operate in a test mode, which forces it to transmit a predefined series of power steps at various reverse link frequencies, and simultaneously tune its receiver to perform measurements (such as RSSI) of the test set's signal at various forward link frequencies and power levels -65 dBm/1.23 MHz to +28 dBm/1.23 MHz 1.25 ms 1.23 MHz brick wall 5 to 20 ms Immediate and RF rise
Measurement range Measurement interval Measurement filter Step size range Trigger sources
Output level setting (to be within 0.1 dB of the final power level) Amplitude switch < 1 ms, typically < 250 s Frequency switch < 1 ms, typically < 500 s within band Measurement accuracy1 (within 45 dB of expected power with < 48 hours warm-up before self alignment initiation) -54 to +28 dBm/1.23 MHz RF source output level range RF in/out port RF out only port < 0.75 dB, typically < 0.3 dB -61 to < -55 dBm/1.23 MHz < 0.85 dB, typically < 0.3 dB -115 dBm/1.23 MHz to -15 dBm/1.23 MHz -115 dBm/1.23 MHz to -5 dBm/1.23 MHz
Composite absolute output level accuracy (< 10 C and < 24 hours from the last alignment, -108 to -15 dBm/1.23 MHz) RF in/out port RF out only port RF output level settling < 1.0 dB, typically 0.65 dB < 1.0 dB, typically 0.65 dB < 1 ms to be within 0.1 dB of the final value
www..com
1. Additional accuracy error when using RF OUT ONLY port is < 0.1 dB.
5
Remove all doubt
Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confidence in your measurements. Agilent offers a wide range of additional expert test and measurement services for your equipment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies' products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada Latin America United States Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand Europe & Middle East Austria Belgium Denmark Finland France Germany 0820 87 44 11 32 (0) 2 404 93 40 45 70 13 15 15 358 (0) 10 855 2100 0825 010 700*
*0.125 fixed network rates
877 894 4414 305 269 7500 800 829 4444
Agilent Email Updates
www.agilent.com/find/emailupdates Get the latest information on the products and applications you select.
1 800 629 485 800 810 0189 800 938 693 1 800 112 929 81 426 56 7832 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008
Agilent Direct
www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence.
01805 24 6333*
*0.14/minute
Agilent Open
www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development.
Ireland 1890 924 204 Israel 972 3 9288 504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 Switzerland (French) 41 (21) 8113811 (Opt 2) Switzerland (German) 0800 80 53 53 (Opt 1) United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus
Revised: October 24, 2007
cdma2000 is a registered certification mark of the Telecommunications Industry Association. Used under license. Product specifications and descriptions in this document subject to change without notice. (c) Agilent Technologies, Inc. 2006, 2007 Printed in USA, December 6, 2007 5989-6426EN
www..com